Technomed EASYCARE600 A=118.0 a0=0.929 a1=0.166 a2=0.132 pACD=5.14 sf=1.36 A=118.3 A=118.5 242 [68],[69],[78],[92]
[222]: data by Dr.H.Lgel, Sulzbach, Germany
Zeiss CT LUCIA 201P/PY A=118.5 a0=1.53 a1=0.40 a2=0.10 pACD=5.67 sf=1.93 A=119.3 A=119.9 48 [316]
[403]: data by Prof. Dr. Rozsival, Kralove, Czech Republic
Ophtec PC 545 QuadrimaX A=118.0 a0=1.03 a1=0.40 a2=0.10 pACD=5.25 sf=1.47 A=118.4 A=118.6 75 [275],[285]
ZEISS AT TORBI 709 MP A=118.3 a0=1.13 a1=0.40 a2=0.10 pACD=5.37 sf=1.59 A=118.5 A=118.3 48 [372]
Europe / Middle East / Africa [336]: data by Dr. Jean Philippe Seguin, Nevers, France
[347]: data by Dr. Ugur Unsal, Izmir, Turkey
Hanita BunnyLens MF A=118.5 a0=1.22 a1=0.40 a2=0.10 pACD=5.50 sf=1.76 A=118.9 A=119.0 36 [338]
[354]: data by Bausch & Lomb
M.Prskavec and M.Wirtitsch, Villach, Austria
[346]: data by Dr. Rudolf Sieben, Aachen, Germany
[334]: data by Dr. Humberto Carreras, Las Palmas, Spain
[281]: data by Medicontur Ltd, Hungary
[6]: based on data by different surgeons from Eye Clinic Kln-Merheim, Germany
[27]: data by Dr.W.Lange, Prien, Germany
All rights reserved. [313]: data by Dr.J.Burger, Deggendorf, Germany
Mexico A-constants provided as a guideline only. Alcon SA60AT (Japan) A=118.4 a0= -1.627 a1=0.305 a2=0.248 pACD=5.64 sf=1.83 A=118.9 A=119.2 501 [41]
The Alcon Monofocal Toric IOL aspheric design delivers a broad refractive landing zone, giving you predictable distance at 20/20 and consistent functional intermediate (20/32 at 66cm) visual acuity. It therefore has a 1:1 ratio with the formulas recommended IOL power outcome. [370]: data by Mr. Azfar W. Chohan, Isle of Man
Hanita SEELENS A=118.6 a0=1.29 a1=0.40 a2=0.10 pACD=5.51 sf=1.73 A=118.9 A=119.2 51 [136]
AMO Tecnis ZMB00 (Hongkong) A=118.8 a0=1.93 a1=0.40 a2=0.10 pACD=6.14 sf=2.35 A=119.7 A=119.7 51 [280]
[21]: data by A.Koletzko,MD, Moers, Germany
[70]: data by HumanOptics AG, Germany
Oculentis LS-412Y A=118.0 a0=0.88 a1=0.40 a2=0.10 pACD=5.07 sf=1.30 A=118.2 A=118.4 44 [106],[323]
[66]: data by Dr.S.Schler, Minden, Germany
[249]: data by Dr.V.I.Apostolov, Amsterdam, The Netherlands
[233]: data by Dr.R.Schmid, Dillingen, Germany
[292]: data by Dr. A. Yildiz, Istanbul, Turkey
Oculentis LU-313T A=118.0 a0=0.97 a1=0.40 a2=0.10 pACD=5.18 sf=1.37 A=118.2 A=118.2 162 [394]
[211]: data by Dr.R.Braig, Holzkirchen, Germany
The single piece models featured a frosted edge and a steeper anterior curvaturechanges that have since been made to the three piece Alcon 8065752134 CENTURION ULTRAVIT 23G ANTERIOR VITRECTOMY PROBE ~ Lot of 6 (No Original Box) Alcon SA60AT Open your eyes to a new world and learn about products and resources for consumers and eyecare professionals alike. HumanOptics Diff-aA(Y) A=118.4 a0=1.28 a1=0.40 a2=0.10 pACD=5.50 sf=1.73 A=118.9 A=119.2 87 [155],[36
The optical design of the To report the clinical performance characteristics of single-piece acrylic Alcon SA30AL and SA60AT intraocular lenses (IOLs). Polytech Polylens AS62 A=118.2 a0= -1.527 a1=0.152 a2=0.237 pACD=5.04 sf=1.28 A=118.2 A=118.4 342 [94],[238],[317]
[332]: data by Dr. Peter Stewart, Maroochydore, QLD Australia
[27]: data by Dr.W.Lange, Prien, Germany
We aspire to lead the world in innovating life-changing vision and eye care products because when people see brilliantly, they live brilliantly 2021 Corporate Responsibility Report Learn more Sight Innovation AMO ReZoom NXG1 A=118.4 a0=0.92 a1=0.40 a2=0.10 pACD=5.20 sf=1.40 A=118.3 A=118.5 113 [63],[131]
Would you like email updates of new search results? [208]: data by Dr.H.Domack, Schweinfurt, Germany
[393]: data by Dres Becker, Cordero, Maruniak. bos8877luckshttp://bro138.epizy.com/sky77liveslot168hoki99slot5000idnggvegas4dlemacauslotsggvegas88gen777gaskan88ligaciputraagen338maxwin138javaplay88138 slotbet88slotokaisar138hoki138big77infini88kaisar88nuke gaming slotdragon77big77.idcuan138emas138jackpot138https://www.infini88.id/bonanza88baccarat onlinepokersbobettangkasnetslot pragmaticslot88slot danapokerseriautowin88warungtotojudi online4d slot, Lorem ipsum dolor sit amet, consectetur adipisicing elit, sed do eiusmod tempor incididunt ut labore et dolore magna aliqua. iMedSales is a premier surgical and medical equipment supplier that provides a combina Philips L12-5 Linear Probe - 50mm Compact 256-Element Transducer ~ BRAND NEW ~, Storz Hopkins 27015B 30 Degree Cystoscope Good Clean Image, STRYKER 502-427-030 VIDEO ARTHROSCOPE, 4MM 30 DEG, Conmed Linvatec E9010 CoolFlex E9000 High Speed Drill, Karl Storz 22220154 HD Image 1 H3M Microscope, Dr 3-Chip Camera, Karl Storz 20132020 Xenon Light Source 175 Watt ~ Lamp only has 13 hours, Medical Blue Nitrile Gloves Latex Free, Powder Free, Textured, Disposable, Non-Sterile 100 pieces per box of 1. AMO Tecnis Z9000 A=119.0 a0= -0.663 a1=0.117 a2=0.232 pACD=5.71 sf=1.93 A=119.2 A=119.5 494 [19],[27],[28],[82],[114]
[383]: data by Dr. KIM Jun Heon, Seoul, South Korea
(+): Identical IOL with same constants available from other source, click for details. Setting: AMO Tecnis ZM900 A=119.0 a0=1.60 a1=0.40 a2=0.10 pACD=5.89 sf=2.21 A=119.8 A=120.5 118 [71],[72],[77],[79]
[41]: data by C.Suto, MD, Tokyo, Japan
When did Woodrow Wilson marry his first wife? ULIB Home
[254]: data by Dr.U.nsal, Izmir, Turkey
Polytech Polylens A60 A=118.0 a0=0.85 a1=0.40 a2=0.10 pACD=5.05 sf=1.31 A=118.2 A=118.6 57 [35]
[383]: data by Dr. KIM Jun Heon, Seoul, South Korea
Based on 2 large datasets (1: N = 888 Hoya Vivinex aberration-correcting and 2: N = 822 Alcon SA60AT spherical lens) a classical formula constant optimisation has been performed for the Hoffer Q, Holladay 1, Haigis and Castrop formulae, to minimise the root mean squared (rms) PE (situation A). [381]: data by Dr. KIM Joung Mok, Seoul, South Korea
Medicontur Z-Flex 690 A=118.0 a0= 0.237 a1=0.315 a2=0.138 pACD=5.05 sf=1.28 A=118.2 A=118.3 437 [281]
[119]: data by Prof.Dr.R.Welt, Ludwigshafen, Germany
K.Gerstmeyer MD(Minden, Germany),
[384]: data by Prof. JOO Choun-Ki, Seaoul, South Korea
[376]: data by Mr. Mark Cahill, Dublin, Ireland
[340]: data by Dr. Hanemoto, Mito, Japan
[362]: data by Dipl.med.Fischer, Dr.Henjes, Schwedt, Torgau, Germany
A private practice in Marshalltown, Iowa, USA. [395]: data by Dr. M. Czubak, Warszawa, Poland
[324]: data by Marc Phelan MD, Abilene, TX, USA
[261]: data by Dr. E.Mertens, Eersel, The Netherlands
[19]: data by K.Gerstmeyer, MD, Minden, Germany
J Cataract Refract Surg.
[227]: data by PD Dr.J.Weindler, Esslingen, Germany
C.Grenauer-Kloevekorn, Halle, Germany
[286]: data by Dr. Martin Drner, Bocholt, Germany
[268]: data by Dr. F.Kesteloot, Kortrijk, Belgium
[296]: data by Dr. A. Kerenyi, Budapest, Hungary
Nidek Nex-Acri AA A=119.5 a0=1.56 a1=0.40 a2=0.10 pACD=5.73 sf=1.95 A=119.2 A=119.7 167 [18]
Rayner Superflex 620 H A=118.0 a0=1.20 a1=0.40 a2=0.10 pACD=5.39 sf=1.64 A=118.8 A=119.0 144 [81],[88]
Kaschube, Linares, Winkgen, Ldenscheid, Germany
[103]: data by Prof.Dr.U.Mester, Sulzbach, Germany
Alcon AcrySof MA60BM A=118.9 a0=1.443 a1=0.077 a2=0.163 pACD=6.08 sf=2.33 A=119.8 A=120.4 271 [2],[10],[18]
[251]: data by Dr.T.Hanemoto, Mito, Japan
[369]: data by Biotech Vision, India and Moss Vision, UK
[296]: data by Dr. A. Kerenyi, Budapest, Hungary
The following IOL constants were derived from clinical results with the ZEISS IOLMaster. [72]: data by Dr.W.Htz, Bad Hersfeld, Germany
[79]: data by Dr.F.Goes, Antwerp, Belgium
The website is intended to provide information that might be of interest to users. [132]: data by C.Lackerbauer, Mnchen, Germany
[309]: data by Dr.F.DeWilde, Sint Martens Latem, Belgium
[266]: data by Dr. A. Hhn, Oldenburg, Germany
[110]: data by Dr.H.Wenkel, Coburg, Germany
Last Revision: Oct 31, 2016 WH, (*): A-const taken from: Mohr A, Hengerer F, Eckardt C: Retropupillare Fixation der Irisklauenlinse bei Aphakie. The Open University is incorporated by Royal Charter (RC 000391), an exempt charity in England & Wales and a charity registered in Scotland . Bausch&Lomb LI61SE Soflex2 A=118.0 a0=1.09 a1=0.40 a2=0.10 pACD=5.31 sf=1.52 A=118.5 A=118.8 98 [25]
There is no single IOL that is best for every person. All three lenses have different strategies for compensating for spherical aberration. Eyeol LW5752R A=118.0 a0=1.27 a1=0.40 a2=0.10 pACD=5.46 sf=1.71 A=118.9 A=119.2 61 [343]
OII Aqua-Sense III-130 A=118.2 a0=1.02 a1=0.40 a2=0.10 pACD=5.23 sf=1.45 A=118.4 A=118.6 30 [146]
Curamed SA700 A=118.8 a0=1.31 a1=0.40 a2=0.10 pACD=5.52 sf=1.75 A=118.8 A=118.9 114 [25]
[394]: data by Dres Scherer,Lojewsky. [200]: data by Dr.D.Klaessen, Eichsttt, Germany
Oculentis LS-313 MF30 A=118.0 a0=0.95 a1=0.40 a2=0.10 pACD=5.21 sf=1.47 A=118.5 A=118.6 239 [259],[304],[311],[312],[327]
Maxwell A, Suryakumar R. Long-term effectiveness and safety of a three-piece acrylic hydrophobic intraocular lens modified with hydroxyethyl-methacrylate: an open-label, 3-year follow-up study. In verbal descriptions, IOL optic centration was assessed to be within 0.4 mm of perfect in 75 (95%) of 79 cases. [105]: data by Wavelight AG, Berlin, Germany
[67]: data by Dr.B.Harrisburg, Sidney, Australia
[299]: data by Cataract Service, Moorfields Eye Hospital, London, UK
[73]: data by Dr.G.Felgemacher, Brilon, Germany
Tekia Tek-Lens Mod.614 A=118.0 a0=0.85 a1=0.40 a2=0.10 pACD=5.05 sf=1.31 A=118.2 A=118.6 57 [35]
2004 Jun;88(6):746-9. doi: 10.1136/bjo.2003.037663. [260]: data by D.Taglia MD, Munster, IN, USA
DO NOT re-sterilize the Clareon IOL by any method.
Moss Vision Optiflex MO/FNY-D003 A=118.0 a0=0.81 a1=0.40 a2=0.10 pACD=4.98 sf=1.19 A=117.9 A=118.1 49 [369]
[47]: data by Dr.Y.Walpuski, Kiel, Germany
White (MA) or Yellow spherical (SN) with aspheric Optic One piece Hydrophobic IOL, extended dioptric Range to +40.0 D. IOLS
[2]: results based on data by
Please use at your own discretion. [3]: based on data by J.L.Parisi MD, Clemson SC, USA and Harvey K.Minatoya MD, Honolulu, HI, USA
Please note also: the constant a0 of the Haigis formula may be positive as well as negative ! Alcon AcrySof IQ ReSTOR Multifocal IOL models (Models SN6AD1, SN6AD3, MN6AD1), this IOL (Model SV25T0) provides an alternate option for clinicians to offer patients with the near add power of +2.5 D, with optimal vision at 53 cm and greater distance dominance in the energy distribution between near and far. Lehmann R, Maxwell A, Lubeck DM, Fong R, Walters TR, Fakadej A. Domilens Nidek Nex-Acri N4-18YG AKTIS AA A=118.4 a0=1.34 a1=0.40 a2=0.10 pACD=5.51 sf=1.73 A=118.9 A=119.2 57 [257]
[226]: data by Dr.T.Neuhann, Mnchen, Germany
Lerche, Hamburg, Germany
Description. than Caucasian eyes. VSY Acriva UD Reviol MF 613 A=118.0 a0=0.65 a1=0.40 a2=0.10 pACD=4.93 sf=1.20 A=118.1 A=118.3 48 [255]
Eyecryl ACTV DIYHS600ROH (*****) A=118.0 a0=0.56 a1=0.40 a2=0.10 pACD=4.67 sf=0.93 A=117.6 A=117.8 57 [349]
[358]: data by Prof. Dr. I.Pallikaris, Iraklion, Crete
[205]: data by Prof.Dr.J.Alio, Alicante, Spain
[307]: data by Prof.Dr.P.Rieck, Berlin, Germany
Ocular Implants, Lens Implants, and Devices More. Alcon, a Novartis Division.
[65]: data by Dr.M.V.Partyngl, Wil, Switzerland
. Constants are given without any legal responsibility ! [228]: data by Dr.R.v.Horenbeeck, Antwerpen, Belgium
ZEISS AT LISA 801 (Acri.LISA 376D) A=117.7 a0=0.82 a1=0.40 a2=0.10 pACD=5.06 sf=1.26 A=118.1 A=118.0 48 [231]
Canon AQ-110NV (Japan) A=119.0 a0=0.107 a1=0.262 a2=0.189 pACD=5.92 sf=2.10 A=119.4 A=119.8 479 [26]
[164]: data by Dr.U.Berens, Karlsruhe, Germany
Biotech Vision Eyecryl Plus ASHFY600 A=118.5 a0=1.20 a1=0.40 a2=0.10 pACD=5.43 sf=1.63 A=118.6 A=118.9 49 [369]
Das KK, Werner L, Collins S, Hong X. [206]: data by Dr.J.L.Rodriguez Prats, Alicante, Spain
[278]: data by Jason Jacobs, MD, Denver, CO, USA
Zemaitiene R, Speckauskas M, Glebauskiene B, Jasinskas V. Kugelberg M, Wejde G, Jayaram H, Zetterstrm C. J Cataract Refract Surg. Corneal Quatrix A=119.6 a0=1.91 a1=0.40 a2=0.10 pACD=6.04 sf=2.28 A=119.8 A=120.4 73 [132],[147]
[119]: data by Prof.Dr.R.Welt, Ludwigshafen, Germany
[366]: data by Dr. Tomas Prusenowsky, Zlin, Czech Republic
[78]: data by Prof.Dr.Th.Neuhann, Mnchen, Germany
[368]: data by Dr. Martin Bechmann, Mnchen, Germany
[399]: data by Afidera GmbH, Berlin, Germany
i-Medical i-Soft A=118.8 a0=1.24 a1=0.40 a2=0.10 pACD=5.44 sf=1.64 A=118.6 A=119.0 107 [128]
Europe / Middle East / Africa This service located at Oisto.com is a website owned and operated by Oisto. [39]: data by Ronald R.Holweger, MD, Hays, KS, USA
[Comparison of postoperative results after implantation of hydrophilic acrylic or hydrophobic acrylic intraocular lens: data of one-year prospective clinical study]. ATTENTION: Refer to the Directions for Use labeling for a complete list of indications, warnings and precautions.REFERENCES: https://professional.myalcon.com/cataract-surgery/intraocular-lens/acrysof-iq-monofocal-iol/. We welcome your input for improving our users Experience. [375]: data by Dr. J. Juranek, Mhldorf am Inn, Deutschland
Clipboard, Search History, and several other advanced features are temporarily unavailable. AMO PS60 ANB A=116.7 a0=1.15 a1=0.40 a2=0.10 pACD=5.46 sf=1.65 A=118.7 A=118.9 18 [2]
Ophtec PC 620 Luna - discontinued A=119.0 a0=1.71 a1=0.40 a2=0.10 pACD=5.91 sf=2.13 A=119.6 A=120.0 52 [247]
[211]: data by Dr.R.Braig, Holzkirchen, Germany
We use cookies to ensure that we give you the best experience on our website. MIRA HI-MY (****) A=120.0 a0=4.19 a1=0.40 a2=0.10 pACD=9.61 sf=5.65 A=122.2 A=118.9 17 [256]
[122]: data by Dr.A.Petzold, Leipzig, Germany
[311]: data by Dr.T.Kohm, Karlsruhe, Germany
Find out more at MyAlcon today! ) Keith A.Gross MD, New Zealand,
XP-55 A=118.0 a0=0.272 a1=0.245 a2=0.154 pACD=5.25 sf=1.48 A=118.5 A=118.8 259 [2],[18]
[395]: data by Dr. M. Czubak, Warszawa, Poland
[322]: data by P.Roholt MD, Canton OH, USA
[243]: data by Jorge Buonsanti MD / Mariana Fuentes MD, Buenos Aires, Argentina
ZEISS CT 47LC (Acri.Lyc 47LC) A=117.7 a0=0.308 a1=0.300 a2=0.133 pACD=4.96 sf=1.17 A=117.9 A=118.0 192 [213],[214]
[139]: data by Dr.C.Deman, Koksijde, Belgium
The biconvex optic and supporting haptics consist of a high refractive index soft acrylic material. [45]: data by Dr.S.Janssen, Belgium
[245]: data by Dr.Klaus Ellendorff, Lneburg, Germany
[101]: data by Peter C.Campanella, MD, Sinking Spring, PA, USA
[401]: data by Dr. F. Kretz, Rheine, Grmany
[328]: data by Dres. [381]: data by Dr. KIM Joung Mok, Seoul, South Korea
Hoya AF-1 VA-60BB A=118.7 a0=1.24 a1=0.40 a2=0.10 pACD=5.43 sf=1.65 A=118.7 A=118.9 50 [81]
[263]: data by Dr. J.-M. Bosq, Nantes, France
[390]: data by Dr. C. Kloos & Dr. B. Pleisch, Augenzentrum Wil, Switzerland
[44]: data by E.Hurst, MD, USA
[117]: data by Dr.M.Theophil, Brandenburg, Germany
[12]: data by E.Frieling, MD,Mnchen, Germany
Eyeol EYE-DIFF Plus A=118.0 a0=0.93 a1=0.40 a2=0.10 pACD=5.15 sf=1.39 A=118.4 A=118.6 75 [243],[371]
[15]: data by J.Chvojkova, MD, Prague, Czech Republic
[34]: data by D.Rotmann, MD, Nordhorn, Germany
SA60AT SA60WF SN60WF CCA0T0 CNA0T0 CZ70BD ACU0T0 SY60WF Packaged service / item; no separate payment made V2630 Anterior chamber intraocular lens MTA3U0 MTA4U0 MTA5U0 Packaged service / item; no separate payment made V2787 Astigmatism-correcting function of intraocular lens (covered conventional IOL component may be reported as V2632) SA6AT3 . The site is secure.
Graefes Arch Clin Exp Ophthalmol. [239]: data by Thomas L. Beardsley, MD, Ashley, NC, USA
[386]: data by Dr. CHO Beom Jin, Incheon, South Korea
[99]: data by IOLtech
Hoya UV55SB (Japan)(**) A=117.6 a0=1.38 a1=0.40 a2=0.10 pACD=5.57 sf=1.68 A=118.6 A=118.5 109 [80]
Aaren Scientific EC-1YH PAL A=117.8 a0=1.20 a1=0.40 a2=0.10 pACD=5.40 sf=1.61 A=118.7 A=118.8 100 [263],[290]
2015 Jun;253(6):973-7. doi: 10.1007/s00417-015-3029-8. [319]: data by Dr. Antoni Salvador Playa, Terrassa, Spain
Medicem WIOL-CF a0=2.36 a1=0.40 a2=0.10 pACD=6.55 sf=2.79 A=120.6 A=121.6 179 [358],[359],[366]
[57]: data by E.Viteri, MD, Ecuador
[338]: data by Dr. Ruth Lapid-Gortzak, Driebergen, The Netherlands
[80]: data by Dr. Kazuya Yamagishi, Japan
Unauthorized use of these marks is strictly prohibited. [392]: data by Dr. M. Gamringer-Kroher, Weiden, Deutschland
Optic glistenings were observed in 11 (11%) of 100 SA30AL cases at a mean follow-up of 8 months. [363]: data by Prof.Dr. Polytech Polylens H11(Y11) (+) A=118.3 a0= -0.037 a1=0.032 a2=0.214 pACD=5.67 sf=1.90 A=119.1 A=119.3 253 [287],[361],[220],[317]
For the Clareon Aspheric Toric IOLs, rotation can reduce astigmatic correction; if necessary lens repositioning should occur as early as possible prior to lens encapsulation. Alcon Toric SN6AT(2-9) A=119.0 a0= -0.323 a1=0.213 a2=0.208 pACD=5.81 sf=1.98 A=119.2 A=119.3 584 [55],[302]
Werner L, Thatthamla I, Ong M, et al. Curamed ASPHERICx HD600YA A=118.4 a0=1.27 a1=0.40 a2=0.10 pACD=5.44 sf=1.68 A=118.8 A=118.9 53 [218]
[165]: data by Dr.R.Friedrich, Meiningen, Germany
[220]: data by Dr.Georg Leuschner, Garbsen, Germany
Uruguay [309]: data by Dr.F.DeWilde, Sint Martens Latem, Belgium
[60]: data by V.Centurion, MD, Brazil
A private practice in Marshalltown, Iowa, USA. 2020 - Slitlamp photographs showed that 52 (98%) of 53 IOLs were in the same position 9 months after surgery as on the day of surgery. Visiontech Medical Optics Liteflex A A=118.2 a0=1.21 a1=0.40 a2=0.10 pACD=5.36 sf=1.57 A=118.6 A=118.7 48 [276]
eyeonics AT-45 crystalens A=119.24 a0= 0.773 a1=0.222 a2=0.160 pACD=5.86 sf=2.06 A=119.3 A=119.4 236 [83]
[26]: data by K.Ichikawa, MD, Aichi, Japan
[361]: data by Dr. Thomas Hamacher, Starnberg, Germany
[171]: data by Dr.S.Sorokanich, Clarks Summit, PA, USA
ZEISS AT LISA tri839 MP A=118.3 a0= -1.477 a1=0.058 a2=0.262 pACD=5.48 sf=1.72 A=118.9 A=119.0 237 [297],[404]
1stQ Basis Z hydrophob (+) A=118.0 a0=1.32 a1=0.40 a2=0.10 pACD=5.46 sf=1.70 A=118.9 A=119.2 97 [271]